test directorySearch Results
Click here to search again.
Test Name:
Phencyclidine, by GC/MS, Serum or Plasma
- SBMF No:
44965 - Performance Lab Name:
ARUP Laboratories - Test Mnemonic:
PCP SP - ABN:
Not required - CPT Code:
83992 - Ref Lab Test No:
0091571 - Also Known As:
Angel Dust
PCP
Phenylcyclohexylpeperidine - Spec Type:
Plasma - Spec Container:
Gray top (fluoride/oxalate), green top (sodium heparin), or lavender top (EDTA) tube - Alt Spec Type:
Serum - Alt Spec Container:
Red top (serum) tube - Pref Vol:
4.0 mL - Min Vol:
2.0 mL - Fasting:
No - Spec Collect:
Do not use gel-barrier tubes for specimen collection
Routine venipuncture - Spec Process:
Serum sample tubes, clot 30 minutes
Promptly centrifuge 15 minutes
Immediately transfer serum or plasma to separate plastic tube - Spec Store Transport:
Refrigerated - Spec Stability:
After separation from cells:
1 week room temperature (20-30°C)
2 weeks refrigerated (2-8°C)
3 years frozen (-20°C) – Do not use frost-free units that undergo repeated freeze/thaw cycles - Spec Reject:
Sample exposed to repeated freeze/thaw cycles
Sample collected in gel-barrier tube
Sample collected in light blue top (sodium citrate) tube - Spec Remarks:
Remove serum/plasma from cells ASAP - Methodology:
Quantitative Gas Chromatography-Mass Spectrometry (GC/MS) - Reference Range:
By reportInterpretive Data:
Drugs covered: phencyclidine (PCP).
Positive cutoff: 10 ng/mLThe concentration value must be greater than or equal to the cutoff to be reported as positive. Interpretive questions should be directed to the laboratory.
For medical purposes only; not valid for forensic use.
- Day Run:
Mon - Time Reported:
2-8 days - Test Type:
TOX-TDM